User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: ME-30E/U DATE: 16-Nov-88 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 100% NUMBER OF TESTS: 21 NUMBER OF LINES: 53 CONFIGURATION: Fluke 5100B ()B ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Turn the Test Instrument Power On and Ensure that the 1.001 DISP Test Instrument has warmed up. 1.002 ASK- F 1.003 HEAD {Volts AC and Linearity Tests} 1.004 DISP Connect the BNC connector from the 7405A AC Divider to 1.004 DISP the INPUT of the Test Instrument using a 1.004 DISP 'BNC Female to Double Banana Adapter'. 1.005 5100 0.001 0.001V 2/ 400H 2W 2.001 5100 0.003 0.003V 2/ 400H 2W 3.001 5100 0.01 0.01V 2/ 400H 2W 4.001 DISP Remove the 7405A AC Divider from the INPUT to the 4.001 DISP Test Instrument. 4.002 DISP Connect the 36" Double Banana to Double Banana Cable 4.002 DISP to the Test Instrument INPUT, and ensure 4.002 DISP correct polarization. 4.003 5100 0.03 0.03V 2/ 400H 2W 5.001 5100 0.1 0.1V 2/ 400H 2W 6.001 5100 0.3 0.3V 2/ 400H 2W 7.001 5100 1 1V 2/ 400H 2W 8.001 5100 3 3V 2/ 400H 2W 9.001 5100 10 3V 2/ 400H 2W 10.001 5100 10 5V 2/ 400H 2W 11.001 5100 10 10V 2/ 400H 2W 12.001 5100 30 30V 2/ 400H 2W 13.001 5100 100 100V 2/ 400H 2W 14.001 5100 300 300V 2/ 400H 2W 15.001 HEAD {Frequency Response Tests} 15.002 DISP Disconnect the Double Banana cable from the Test 15.002 DISP Instrument. 15.003 DISP Connect the 'BNC Female to Double Banana Adapter' to the 15.003 DISP Test Instrument INPUT. 15.003 DISP Connect the 50Ohm Feedthrough to that adapter. 15.003 DISP Connect the 36" Coaxial Cable to the 50Ohm Feedthrough. 15.004 WIDE 0.3 0.3V 2/ 400H N 2W 15.005 MEMC 0.3 .3V 20% 400H_Reference 16.001 WIDE 0.3 0.3V 5/ 10H C 2W 17.001 WIDE 0.3 0.3V 2/ 20H C 2W 18.001 WIDE 0.3 0.3V 2/ 1MH C 2W 19.001 WIDE 0.3 0.3V 3/ 2MH C 2W 20.001 WIDE 0.3 0.3V 5/ 4MH C 2W 21.001 END